100S Reference Manual
SETUP
(
S - M
M
LEC =
)
x 1,000,000
S = Standard length
M = Measured length
In the DRO mode.
Place a standard of known length on the table. Make sure it’s parallel
with the table’s travel.
Put the readout in the absolute display mode (
ABS/INCR key).
Using an edge finder or dial indicator, locate one end of the standard.
Press the
ZERO key twice for the axis you are measuring. “0” should
appear on the display.
Move the table along the axis until the edge finder or dial indicator
reaches the opposite end of the standard.
Write down the length that is shown on the readout display for
that axis.
Use the formula below to calculate the LEC for the axis you just
measured. Enter the result in the LEC screen (previous page).
Note
: If the measured length is greater than the standard length the
LEC value will be negative.
How to Determine the Linear Error Compensation
Follow this procedure for each axis:
Formula
Example
If the length of the
standard you used is
10” and the measured
length along the X-
axis is 9.995”, then
the LEC for the X-
axis is 500 parts per
million (PPM).
(
10 - 9.995
9.995
LEC =
)
x 1,000,000
LEC =
500 PPM
(rounded to the nearest
whole number)